FDA Adverse Event Malfunction Summary report: N

DIMENSION EXL WITH LM

MDR report key: 3866898 · Received June 11, 2014

Report

Report Number
1226181-2014-00315
Event Type
Malfunction
Date Received
June 11, 2014
Date of Event
May 17, 2014
Report Date
May 17, 2014
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC
Product Code
JJE
PMA / PMN Number
K130276
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACTED THE SIEMENS TECHNICAL SOLUTIONS CENTER (TSC). THE CUSTOMER STATED THAT A LEAK WAS DISCOVERED IN THE SAMPLE PROBE ASSEMBLY. A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER SITE. AFTER EVALUATION OF THE INSTRUMENT AND INSTRUMENT DATA, THE CSE REPLACED THE SAMPLE PROBE AND KINKED TUBING. THE CAUSE OF THE DISCORDANT, FALSELY LOW NA, K AND CL RESULTS WAS A LEAK IN THE SAMPLE PROBE. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

DISCORDANT, FALSELY LOW SODIUM (NA), POTASSIUM (K) AND CHLORIDE (CL) RESULTS WERE OBTAINED ON ONE PATIENT SAMPLE ON A DIMENSION EXL WITH LM INSTRUMENT. THE DISCORDANT RESULTS WERE NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED ON AN ALTERNATE INSTRUMENT, RESULTING HIGHER. THE CORRECTED RESULTS FROM THE ALTERNATE INSTRUMENT WERE REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY LOW NA, K AND CL RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344653 DIMENSION EXL WITH LM CLINICAL CHEMISTRY ANALYZER JJE SIEMENS HEALTHCARE DIAGNOSTICS INC DIMENSION EXL WITH LM

Patients

Seq Age Sex Outcome Treatment
1