FDA Adverse Event Malfunction Summary report: N

DONOR SCREEN HLA CLASS I ANTIBODY DETECTION ASSAY

MDR report key: 3866691 · Received June 3, 2014

Report

Report Number
MW5036492
Event Type
Malfunction
Date Received
June 3, 2014
Date of Event
May 28, 2014
Report Date
June 3, 2014
Manufacturer
GENETIC TESTING INSTITUTE, INC (GTI), GENPROBE LIFECODES
Product Code
MZI
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

AN ABNORMALLY HIGH REACTIVITY RATE FOR HLA CLASS I AND CLASS II ANTIBODIES IN VOLUNTEER BLOOD DONORS PROMPTED AN INVESTIGATION OF THE PRECISION OF THE LIFE-CODES DONOR SCREEN HLA CLASS I AND CLASS II ANTIBODY DETECTION ASSAY (B)(4) ON THE QUICKSTEP PLATFORM. A PRECISION STUDY WAS PERFORMED USING THE STATISTICAL FORMULAE RECOMMENDED IN CLSI EP 15-A2 TO ANALYZE THE QUANTITIVE OD AND THE RESULTS INDICATE THAT THE TEST DOES NOT APPEAR TO BE PERFORMING WITHIN THE PARAMETERS DEFINED IN THE PACKAGE INSERT, SPECIFICALLY THE REPRODUCIBILITY AS EVIDENCED BY THE STANDARD DEVIATION BETWEEN DAY TO DAY RUNS. INTRA-VARIABILITY IS BETTER, BUT STILL DEMONSTRATES GREATER IMPRECISION AT SOME LEVELS THAN EXPECTED PER PACKAGE INSERT. ALTHOUGH THIS IS A QUALITATIVE ASSAY, THE TARGET PRECISION ACCEPTABILITY VALUES WERE BASED ON THE QUALITATIVE OD VALUES PRESENT IN THE LIFECODES MANUFACTURER'S DIRECTION INSERT. IT WAS ALSO OBSERVED THAT THE QUALITY CONTROL DATA GENERATED ON THE QUICKSTEP DOES NOT ASSUME A GAUSSIAN DISTRIBUTION. THE LABORATORY WHERE TESTING IS PERFORMED IS TEMPERATURE CONTROLLED AND MONITORED. MANUFACTURER'S DIRECTION INSERTS ARE STRICTLY FOLLOWED TO ENSURE REAGENTS ARE UTILIZED APPROPRIATELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323689 DONOR SCREEN HLA CLASS I ANTIBODY DETECTION ASSAY DONOR SCREEN HLA CLASS I MZI GENETIC TESTING INSTITUTE, INC (GTI), GENPROBE LIFECODES
323690 DONOR SCREEN HLA CLASS II ANTIBODY DETECTION ASSAY DONOR SCREEN HLA II MZI GENETIC TESTING INSTITUTE, INC (GTI), GENPROBE LIFECODES

Patients

Seq Age Sex Outcome Treatment
1 Other