FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 3865641 · Received June 11, 2014

Report

Report Number
9611451-2014-00499
Event Type
Malfunction
Date Received
June 11, 2014
Report Date
May 13, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE TWO COMPLAINT CHAMBERS WERE RECEIVED AT FISHER & PAYKEL HEALTHCARE (B)(4). ONE CHAMBER WAS FROM LOT 130424, MANUFACTURED 24 APR 2013. THE SECOND CHAMBER WAS FROM LOT 130816, MANUFACTURED 16 AUG 2013. BOTH CHAMBERS WERE VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THE CHAMBER DOMES WERE CRACKED NEAR THE BAFFLE. A LOT CHECK REVEALED NO OTHER SIMILAR COMPLAINTS FOR EITHER LOT NUMBER. CONCLUSION: THE NATURE OF THE CRACKING IS CONSISTENT WITH THE CHAMBERS BEING USED WITH A SIPAP VENTILATOR. THIS TYPE OF VENTILATOR IS CAPABLE OF PRODUCING PRESSURES IN EXCESS OF 80CMH2O. THE INTEGRITY OF THE CHAMBER CAN BE AFFECTED WHEN PRESSURES EXCEED 80CMH2O. EVERY MR290 CHAMBER IS PRESSURE TESTED FOLLOWING THE MANUFACTURING PROCESS TO CHECK FOR ANY LEAKS PRESENT IN THE CHAMBER DOME DUE TO CRACKS AND OTHER CAUSES. IN ADDITION, THE PRESSURE TEST IS FOLLOWED BY A VISUAL INSPECTION OF EACH CHAMBER. ANY CHAMBER WHICH FAILS EITHER OF THESE TESTS IS REJECTED. THE CHAMBERS WOULD HAVE MET THE REQUIRED SPECIFICATION AT THE TIME OF PRODUCTION. OUR USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: USE OF THE MR290 ABOVE THE MAXIMUM OPERATING PRESSURE [8KPA (~80CMH2O)] MAY LEAD TO CRACKING, WATER LEAKAGE AND, ON RARE OCCASIONS COULD LEAD TO A LOSS OF VENTILATION PRESSURE. PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT WATER LEAK OCCURRED FROM TWO MR290 AUTOFEED CHAMBERS. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
344332 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 130424

Patients

Seq Age Sex Outcome Treatment
1