ECHELON LINEAR CUTTER
Report
- Report Number
- 3005075853-2014-03917
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- April 24, 2014
- Report Date
- April 24, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K081146
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). RESULTS; DAMAGED YOKE. THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED WITH THE CLAMPING MECHANISM DAMAGED AND WITHOUT RELOAD PRESENT. NO FUNCTIONAL TEST WAS PERFORMED DUE THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE INTERNAL COMPONENTS AND THE YOKE WAS NOTED TO BE BROKEN. ALTHOUGH NO CONCLUSION COULD BE REACHED AS TO HOW THE YOKE BECAME BROKEN, THIS DAMAGE CAN OCCUR WHEN EXCESSIVE FORCE IS APPLIED TO THE CLOSING TRIGGER WHEN THE JAWS ARE CLAMPED OVER AN EXCESS OF TISSUE. IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4).
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE WOULD NOT FIRE. A SECOND DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340767 | ECHELON LINEAR CUTTER | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | L90L02 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |