FDA Adverse Event Malfunction Summary report: N

ECHELON LINEAR CUTTER

MDR report key: 3865020 · Received June 10, 2014

Report

Report Number
3005075853-2014-03917
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 24, 2014
Report Date
April 24, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS; DAMAGED YOKE. THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED WITH THE CLAMPING MECHANISM DAMAGED AND WITHOUT RELOAD PRESENT. NO FUNCTIONAL TEST WAS PERFORMED DUE THE CONDITION OF THE DEVICE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE INTERNAL COMPONENTS AND THE YOKE WAS NOTED TO BE BROKEN. ALTHOUGH NO CONCLUSION COULD BE REACHED AS TO HOW THE YOKE BECAME BROKEN, THIS DAMAGE CAN OCCUR WHEN EXCESSIVE FORCE IS APPLIED TO THE CLOSING TRIGGER WHEN THE JAWS ARE CLAMPED OVER AN EXCESS OF TISSUE. IT SHOULD BE NOTED THAT A 100% INSPECTIONS TAKES PLACE DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRE SPECIFICATIONS; IN ADDITION, A SAMPLE OF THE BATCH IS INSPECTED AT (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE DEVICE WOULD NOT FIRE. A SECOND DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340767 ECHELON LINEAR CUTTER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA L90L02

Patients

Seq Age Sex Outcome Treatment
1