FDA Adverse Event Injury Summary report: N

CONSULTA CRT-P

MDR report key: 3864925 · Received June 10, 2014

Report

Report Number
9614453-2014-01482
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 1, 2014
Report Date
April 1, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NVZ
PMA / PMN Number
P010015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCTS: 407645 LEAD IMPLANTED: 2013 (B)(6); 3778-60 X 2 PAIN STIM LEAD IMPLANTED: 2007 (B)(6); 37742 NEURO PROGRAMMER IMPLANTED: 2007 (B)(6); 37711 PAIN STIM IPG IMPLANTED: 2007 (B)(6); 429688 LEAD IMPLANTED: 2013 (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH PACING THRESHOLDS WITH INTERMITTENT CAPTURE. ALSO, THE IMPLANTABLE PULSE GENERATOR (IPG) EXPERIENCED UNEXPECTED LONGEVITY. THE LEAD AND DEVICE WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340542 CONSULTA CRT-P PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ IPG MFG SWITZERLAND C4TR01

Patients

Seq Age Sex Outcome Treatment
1 00079 YR Hospitalization| R 407658 LEAD