FDA Adverse Event Malfunction Summary report: N

CONSULTA CRT-D

MDR report key: 3864918 · Received June 10, 2014

Report

Report Number
9614453-2014-01480
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 1, 2014
Report Date
April 2, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
NIK
PMA / PMN Number
P010031
Removal / Correction Number
Z-0111-2011
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY:THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED HIGH VOLTAGE CAPACITOR CHARGE TIME WAS LONGER THAN EXPECTED FOR A DEVICE AT ERI (ELECTIVE REPLACEMENT INDICATOR). ERI WAS TRIGGERED ON (B)(6) 2014. EXCESSIVE CHARGE TIME CAPTURED ON (B)(6) 2014.

Additional Manufacturer Narrative · 1

THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES; HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S. THIS EVENT OCCURRED OUTSIDE THE US AND PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION.  BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXPERIENCED LONG CHARGE TIMES. END OF SERVICE (EOS) WAS REACHED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.