FDA Adverse Event Injury Summary report: N

CAPSURE Z NOVUS

MDR report key: 3864908 · Received June 10, 2014

Report

Report Number
2649622-2014-07455
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 28, 2014
Report Date
April 28, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P850089
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED, AND THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT PRODUCT: SEDR01, IPG, IMPLANTED: (B)(6) 2003. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD IMPEDANCE AND THRESHOLD WERE HIGH. THE RV LEAD WAS CAPPED AND REPLACED. IT WAS FURTHER REPORTED THAT THE ATRIAL LEAD HAD NOISE AND FRACTURED. THE ATRIAL LEAD WAS REMOVED, BUT NO NEW ATRIAL LEAD WAS IMPLANTED IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339480 CAPSURE Z NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5554

Patients

Seq Age Sex Outcome Treatment
1 00057 YR Hospitalization| R 5054, LEAD