CAPSURE Z NOVUS
Report
- Report Number
- 2649622-2014-07455
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- April 28, 2014
- Report Date
- April 28, 2014
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P850089
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, ANALYZED, AND THE DISTAL CONDUCTOR OF THE LEAD DEVELOPED A FRACTURE DUE TO FLEXING WHILE IN VIVO. THE OUTER INSULATION OF THE LEAD DEVELOPED A BREACH DUE TO A DEPRESSION WHILE IN VIVO.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. CONCOMITANT PRODUCT: SEDR01, IPG, IMPLANTED: (B)(6) 2003. (B)(4).
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD IMPEDANCE AND THRESHOLD WERE HIGH. THE RV LEAD WAS CAPPED AND REPLACED. IT WAS FURTHER REPORTED THAT THE ATRIAL LEAD HAD NOISE AND FRACTURED. THE ATRIAL LEAD WAS REMOVED, BUT NO NEW ATRIAL LEAD WAS IMPLANTED IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 339480 | CAPSURE Z NOVUS | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 5554 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00057 YR | Hospitalization| R | 5054, LEAD |