FDA Adverse Event Malfunction Summary report: N

EVERA XT DR

MDR report key: 3864893 · Received June 10, 2014

Report

Report Number
3004209178-2014-11331
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
February 27, 2014
Report Date
March 3, 2014
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE RETURNED DEVICE INDICATED A DAMAGED GROMMET AND A LOOSE/DETACHED SET SCREW.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD)  WAS ATTEMPTED BUT NOT USED DURING IMPLANT PROCEDURE DUE TO A GROMMET PROBLEM. WHEN THE PHYSICIAN WAS INSERTING THE ATRIAL LEAD, THE GROMMET UNSCREWED FROM THE CONNECTOR. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338248 EVERA XT DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO DDBB1D4

Patients

Seq Age Sex Outcome Treatment
1 00074 YR