FDA Adverse Event
Malfunction
Summary report: N
EVERA XT DR
MDR report key: 3864893
·
Received June 10, 2014
Report
- Report Number
- 3004209178-2014-11331
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- February 27, 2014
- Report Date
- March 3, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- LWS
- PMA / PMN Number
- P980016
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE RETURNED DEVICE INDICATED A DAMAGED GROMMET AND A LOOSE/DETACHED SET SCREW.
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Additional Manufacturer Narrative · 1
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS ATTEMPTED BUT NOT USED DURING IMPLANT PROCEDURE DUE TO A GROMMET PROBLEM. WHEN THE PHYSICIAN WAS INSERTING THE ATRIAL LEAD, THE GROMMET UNSCREWED FROM THE CONNECTOR. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 338248 | EVERA XT DR | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | DDBB1D4 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00074 YR |