FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3864817 · Received June 10, 2014

Report

Report Number
2649622-2014-07495
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 26, 2014
Report Date
March 26, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5076-52 LEAD IMPLANTED: (B)(6) 2014. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT ONE DAY AFTER THE IMPLANT OF THE PACING LEAD IT EXHIBITED HIGH IMPEDANCE AND A MICRO LEAD DISLODGEMENT WAS SUSPECTED. THE LEAD WAS REPOSITIONED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338122 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00072 YR Required Intervention ADDR01 IPG