FDA Adverse Event Injury Summary report: N

ATTAIN ABILITY

MDR report key: 3864811 · Received June 10, 2014

Report

Report Number
2649622-2014-07520
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 13, 2014
Report Date
March 14, 2014
Manufacturer
MPRI
Product Code
OJX
PMA / PMN Number
P080006
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 6947-62, LEAD, IMPLANTED: (B)(6) 2014; 5076-52, LEAD, IMPLANTED: (B)(6) 2014. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED STRETCHING AND APPARENT EXPLANT DAMAGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD WAS DIFFICULT TO PLACE DUE TO PATIENT ANATOMY. THE MORNING AFTER IMPLANT, THE LEAD EXHIBITED HIGH THRESHOLDS DUE TO DISLODGEMENT. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338120 ATTAIN ABILITY DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE OJX MPRI 419688

Patients

Seq Age Sex Outcome Treatment
1 00074 YR Hospitalization| R D334TRM ICD