CAPSUREFIX MRI SURESCAN
Report
- Report Number
- 2649622-2014-07541
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 26, 2014
- Report Date
- March 26, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: A2DR01 IMPLANTABLE PULSE GENERATOR (IPG), IMPLANTED: (B)(6) 2013. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE ACTUAL LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY WAS PERFORMED AND NO ANOMALIES WERE FOUND.
IT WAS REPORTED THAT THE PATIENT PRESENTED COMPLAINING OF CHEST WALL PAIN DURING VENTRICULAR PACING THRESHOLD TESTING. INTRA-OPERATIVE INTERROGATION REVEALED NORMAL LIMITS. HIGH OUTPUT PACING PERFORMED ON THE VENTRICULAR LEAD REPRODUCED THE SAME DISCOMFORT. A MICRO-PERFORATION IS SUSPECTED. THE RV LEAD WAS REPOSITIONED FORM A DISTAL SEPTAL POSITION TO AN APICAL POSITION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 337764 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00050 YR | Life Threatening| R | 5086MRI52 LEAD |