FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3864741 · Received June 10, 2014

Report

Report Number
2649622-2014-07541
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 26, 2014
Report Date
March 26, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: A2DR01 IMPLANTABLE PULSE GENERATOR (IPG), IMPLANTED: (B)(6) 2013. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE ACTUAL LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY WAS PERFORMED AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED COMPLAINING OF CHEST WALL PAIN DURING VENTRICULAR PACING THRESHOLD TESTING. INTRA-OPERATIVE INTERROGATION REVEALED NORMAL LIMITS. HIGH OUTPUT PACING PERFORMED ON THE VENTRICULAR LEAD REPRODUCED THE SAME DISCOMFORT. A MICRO-PERFORATION IS SUSPECTED. THE RV LEAD WAS REPOSITIONED FORM A DISTAL SEPTAL POSITION TO AN APICAL POSITION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337764 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI58

Patients

Seq Age Sex Outcome Treatment
1 00050 YR Life Threatening| R 5086MRI52 LEAD