PROTECTA XT CRT-D
Report
- Report Number
- 3004209178-2014-11383
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 31, 2014
- Report Date
- March 31, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NIK
- PMA / PMN Number
- P010031
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 693565, LEAD, IMPLANTED: (B)(6) 2010; 419688, LEAD, IMPLANTED: (B)(6) 2013. (B)(4).
PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED, A ANALYZED, AND NO ANOMALIES WERE FOUND.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCKS. THE RIGHT VENTRICULAR (RV) LEAD HAD AN INCREASE IN PACE/SENSE IMPEDANCES, OVERSENSING, AND EPISODES WITH NOISE. LEAD FRACTURE IS SUSPECTED. IT WAS ALSO NOTED THAT IT COULD POSSIBLY BE A CONNECTION OR HEADER ISSUE. THE LEAD AND DEVICE WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 337714 | PROTECTA XT CRT-D | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO | NIK | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | D314TRG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00063 YR | Hospitalization| R | 383059 LEAD |