FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC ACE

MDR report key: 3864490 · Received June 10, 2014

Report

Report Number
3005075853-2014-03908
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
May 29, 2014
Report Date
June 5, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K042777
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). POTENTIAL REPROCESSING; A NEW BLADE CANNOT BREAK UNLESS IT WAS CRACKED BY PRIOR USE. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RETURNED IN GOOD PHYSICAL CONDITION WITH THE BLADE TIP INTACT AND NOT BROKEN OFF AS REPORTED BY THE CUSTOMER. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON GEN11 AN ALERT SCREEN WAS DISPLAYED. THE DEVICE WAS DISASSEMBLED AND IT WAS FOUND THAT THE BLADE WAS CRACKED AT THE PIN HOLE UNDER THE SHROUD OF THE DEVICE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN ALERT SCREENS, SUCH AS ¿TIGHTEN ASSEMBLY¿, ¿BLADE ERROR DETECTED¿ OR "RELAXED PRESSURE ON BLADE" FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE.

Additional Manufacturer Narrative · 1

(B)(4). AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS SHOULD THE DEVICE BE RETURNED FOR ANALYSIS, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE DEVICE WAS NOT RECEIVED FOR ANALYSIS; THEREFORE, AN INVESTIGATION COULD NOT BE PERFORMED. WE DID NOT RECEIVE A BATCH NUMBER OR LOT NUMBER SO THEREFORE WE WERE UNABLE TO REVIEW THE MANUFACTURING RECORDS.

Description of Event or Problem · 1

IT WAS REPORTED THAT BEFORE AN UNKNOWN PROCEDURE, THE TITANIUM TIP BROKE DURING THE PRE-RUN TEST. THE BROKEN PIECE DID NOT FALL INTO THE PATIENT¿S BODY. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED. ONE DEVICE WILL BE RETURNED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338955 ULTRACISION HARMONIC ACE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE