FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3864408 · Received June 10, 2014

Report

Report Number
3004209178-2014-85531
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
June 5, 2014
Report Date
June 5, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST DUE A PROTRUDED DRIVE SUPPORT DISK. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED CASE AT DISPLAY WINDOW CORNERS, A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND A CRACKED BELT CLIP SLOT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED AND WAS STUCK IN THE REWIND MODE. IT WAS STATED THAT THE DEVICE WAS DROPPED. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP WAS PROTRUDED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341249 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWB

Patients

Seq Age Sex Outcome Treatment
1