FDA Adverse Event Malfunction Summary report: N

XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3864383 · Received June 10, 2014

Report

Report Number
2024168-2014-03715
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
May 16, 2014
Report Date
May 16, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED STENT DISLODGEMENT WAS CONFIRMED. BASED ON VISUAL AND DIMENSIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY OF THE COMPLAINT HANDLING DATABASE REVEALED NO OTHER INCIDENTS FOR STENT DISLODGMENT REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT USING A FEMORAL ARTERY ACCESS APPROACH DURING A PROCEDURE OF THE UNSPECIFIED ARTERY, THE 3.5 X 23 MM XIENCE XPEDITION STENT IMPLANT DISLODGED OFF THE BALLOON WHILE ATTEMPTING TO INSERT THE STENT DELIVERY SYSTEM (SDS) OVER THE GUIDE WIRE. THERE WAS NO PATIENT INTERACTION. THE DEVICE WAS NOT USED. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341176 XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 4010241

Patients

Seq Age Sex Outcome Treatment
1