FDA Adverse Event Malfunction Summary report: N

ELECTRIC PEN DRIVE 60,000 RPM

MDR report key: 3864333 · Received June 10, 2014

Report

Report Number
2520274-2014-11663
Event Type
Malfunction
Date Received
June 10, 2014
Report Date
May 22, 2014
Manufacturer
SYNTHES OBERDORF
Product Code
HWE
PMA / PMN Number
PK043310
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURING LOCATION WAS UNKNOWN. THE DATE OF MANUFACTURE WAS UNKNOWN. THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE DOES NOT FUNCTION WITH THE HAND CONTROL DEVICE. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO A DAMAGED MOTOR ASSEMBLY DUE TO NORMAL USE OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE MANUFACTURER LOCATION WAS DOCUMENTED AS UNKNOWN IN THE INITIAL REPORT. THE LOCATION HAS BEEN UPDATED TO (B)(4). CONTACT OFFICE NAME/ADDRESS HAS BEEN UPDATED ACCORDINGLY TO REFLECT THE CORRECT MANUFACTURING FACILITY. DEVICE MANUFACTURE DATE: THE DEVICE MANUFACTURE DATE WAS DOCUMENTED AS UNKNOWN IN THE INITIAL REPORT. THE DEVICE MANUFACTURE DATE HAS BEEN UPDATED AS OCT 6, 2010. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ELECTRIC PEN DRIVE DEVICE WOULD NOT "ACTIVATE WITH THE HANDLE". THE EVENT WAS NOT REPORTED TO HAVE OCCURRED DURING SURGERY. IT WAS UNKNOWN TO THE REPORTER IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. NO INJURIES OR MEDICAL INTERVENTION WERE REPORTED. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339499 ELECTRIC PEN DRIVE 60,000 RPM INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT HWE SYNTHES OBERDORF 4879

Patients

Seq Age Sex Outcome Treatment
1