RX ACCULINK CAROTID STENT SYSTEM
Report
- Report Number
- 2024168-2014-03716
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- May 16, 2014
- Report Date
- May 16, 2014
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- NIM
- PMA / PMN Number
- P040012
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION AND THE REPORTED HANDLE DAMAGE WAS CONFIRMED. BASED ON VISUAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY OF THE COMPLAINT HANDLING DATABASE REVEALED NO OTHER INCIDENTS REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT AFTER DEVICE PREPARATION AND WHILE ATTEMPTING TO BACKLOAD THE DEVICE ONTO THE GUIDE WIRE THE SHAFT AND SLIDER HANDLE WAS NOTED TO HAVE SEPARATED INTO TWO PIECES. THE DEVICE WAS NOT USED. THERE WAS NO PATIENT INTERACTION. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 339498 | RX ACCULINK CAROTID STENT SYSTEM | CAROTID STENT SYSTEM | NIM | AV-TEMECULA-CT | 4013161 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |