PROGRASP FORCEPS INSTRUMENT INSTRUMENT
Report
- Report Number
- 2955842-2014-03563
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- May 13, 2014
- Report Date
- May 13, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT THE PITCH CABLE WAS FRAYED AT THE DISTAL CLEVIS HUB. NO DAMAGE WAS FOUND AT THE CLEVIS. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THE FOLLOWING ADDITIONAL DAMAGES TO THE INSTRUMENT: THE PITCH CABLE WAS FOUND LOOSE AT THE DISTAL CLEVIS HUB. BOTH CABLE CRIMPS REMAINED IN THE CLEVIS AND FOUND FRAYED AT THE DISTAL CLEVIS HUB AREA. THE HOUSING WAS REMOVED AND THE PITCH CABLE WAS FOUND LOOSE AT THE CLAMPING PULLEY, BUT NO LOOSE CLAMPING PULLEY SCREW WAS FOUND. THE CLAMPING PULLEY WAS CORRODED/CONTAMINATED. RESIDUE WAS FOUND AT THE CABLE RAILS AREA ON THE CLAMPING PULLEY. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE WAS LIKELY DUE TO IMPROPER CLEANING. THERE WAS AN INDENTATION AT THE INNER EDGE OF THE DISTAL PULLEY AND VISIBLE SCRATCHES ON THE SURFACE OF THE PULLEY. THE EVIDENCE WAS INCONCLUSIVE, BUT DAMAGE WAS LIKELY DUE TO MISHANDLING/MISUSE. THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS EXHIBITING LIGHT MATERIAL REMOVAL AND A ROUGH SURFACE FINISH. THE SCRATCHES WERE SHORT IN LENGTH AND NOT AXIALLY ALIGNED WITH THE MAIN TUBE. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST® INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE REPROCESSING INSTRUCTIONS UNDER CLEANING, DISINFECTION, AND STERILIZATION GENERAL INFORMATION SECTION SPECIFICALLY STATES: WARNING: READ ALL INSTRUCTIONS CAREFULLY. FAILURE TO PROPERLY FOLLOW INSTRUCTIONS MAY CAUSE IMPROPER FUNCTIONING OF THE DEVICE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE CABLE DAMAGE AND/OR THE TUBE ABRASIONS, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT THE PROGRASP FORCEPS INSTRUMENT WIRE WAS NOTED TO BE BROKEN PRIOR TO STARTING A DA VINCI SURGICAL PROCEDURE. NO FRAGMENTS FELL INTO THE PATIENT, NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 337623 | PROGRASP FORCEPS INSTRUMENT INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420093-08 | M10131008 976 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |