FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3864248 · Received June 10, 2014

Report

Report Number
2955842-2014-03560
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 23, 2014
Report Date
May 13, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS WAS UNABLE TO REPLICATE THE CUSTOMER REPORTED COMPLAINT. FAILURE ANALYSIS INSPECTED THE INSTRUMENT UNDER DIGITAL MICROSCOPE AND NO DAMAGED WIRES WERE FOUND. FAILURE ANALYSIS ALSO OBSERVED THAT THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE TUBE ABRASIONS, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE MARYLAND BIPOLAR FORCEPS INSTRUMENT WIRES AT THE WRIST WERE NOTED TO BE KINKED PRIOR TO STARTING A DA VINCI HYSTERECTOMY PROCEDURE. NO FRAGMENTS FELL INTO THE PATIENT, NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338157 MARYLAND BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420172-07 N10131109 931

Patients

Seq Age Sex Outcome Treatment
1