FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 3864246 · Received June 10, 2014

Report

Report Number
2955842-2014-03558
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
March 24, 2014
Report Date
May 13, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K061260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS FOUND THAT THE PITCH CABLE WAS FRAYED AT THE PROXIMAL CLEVIS HUB. THE CLEVIS DID NOT EXHIBIT ANY WEAR. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE DAMAGED CABLE, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PK DISSECTING FORCEPS INSTRUMENT WOULD NOT RESPOND TO THE SURGEON'S MOVEMENTS IN THE CONSOLE DURING A DA VINCI HYSTERECTOMY PROCEDURE. SEVERAL TROUBLESHOOTING MANEUVERS WERE DONE INCLUDING RESEATING THE INSTRUMENT AND THE CASSETTE DRAPE TO THE INSTRUMENT ARM. A BACKUP INSTRUMENT HAD TO BE OPENED AND THE PLANNED PROCEDURE WAS COMPLETED. NO FRAGMENTS FELL INTO THE PATIENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338547 PK DISSECTING FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420227-03 N10131209 950

Patients

Seq Age Sex Outcome Treatment
1