FDA Adverse Event Malfunction Summary report: N

LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT

MDR report key: 3864140 · Received June 10, 2014

Report

Report Number
2955842-2014-03556
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
May 7, 2014
Report Date
May 14, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT THE GRIP CABLE WAS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND WAS UNDAMAGED. THE CABLE SEGMENT STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. FAILURE ANALYSIS ALSO OBSERVED SCRATCHES ON THE SURFACE OF THE DISTAL PULLEY. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE DAMAGED CABLE, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LARGE SUTURECUT NEEDLE DRIVER INSTRUMENT CABLE THAT CONTROLS SOME OF THE MECHANICAL CAPABILITIES OF THE INSTRUMENT COMPLETELY SNAPPED DURING A DA VINCI GASTRECTOMY PROCEDURE. NO FRAGMENTS FELL INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338427 LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420296-02 M10130723 840

Patients

Seq Age Sex Outcome Treatment
1 69 YR