LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT
Report
- Report Number
- 2955842-2014-03556
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- May 7, 2014
- Report Date
- May 14, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT THE GRIP CABLE WAS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND WAS UNDAMAGED. THE CABLE SEGMENT STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. FAILURE ANALYSIS ALSO OBSERVED SCRATCHES ON THE SURFACE OF THE DISTAL PULLEY. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE DAMAGED CABLE, COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT THE LARGE SUTURECUT NEEDLE DRIVER INSTRUMENT CABLE THAT CONTROLS SOME OF THE MECHANICAL CAPABILITIES OF THE INSTRUMENT COMPLETELY SNAPPED DURING A DA VINCI GASTRECTOMY PROCEDURE. NO FRAGMENTS FELL INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 338427 | LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420296-02 | M10130723 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR |