FDA Adverse Event Malfunction Summary report: N

DOUBLE FENESTRATED GRASPER INSTRUMENT

MDR report key: 3863978 · Received June 10, 2014

Report

Report Number
2955842-2014-03548
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
May 5, 2014
Report Date
May 15, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND A LOOSE PITCH CABLE DUE TO ANOTHER PITCH CABLE AT THE TIP WAS BROKEN. ADDITIONAL OBSERVATION NOR REPORTED WAS A BROKEN PITCH UP CABLE AT THE DISTAL CLEVIS HUB. THE BROKEN CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE BROKEN CABLE FOUND DURING FAILURE ANALYSIS COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT IN CENTRAL PROCESSING, A LOOSE CABLE WAS IDENTIFIED ON A DOUBLE FENESTRATED GRASPER INSTRUMENT. THERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT. NOTHING REPORTEDLY FELL INTO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339054 DOUBLE FENESTRATED GRASPER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420189-09 M10131108 851

Patients

Seq Age Sex Outcome Treatment
1