- 1000165971-2014-00335
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- May 16, 2014
- Report Date
- May 20, 2014
- Manufacturer
- SORIN GROUP ITALIA S.R.L. - CRM FACILITY
- Product Code
- NVZ
- PMA / PMN Number
- PP950029
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
REPORTEDLY, ON (B)(6) 2014, DURING A REGULAR FOLLOW-UP, THE PACEMAKER REACHED RECOMMENDED REPLACEMENT TIME (RRT) AND BATTERY IMPEDANCE WAS AT 10KOHMS. ACCORDING TO THE PATIENT, THE PATIENT FELT SOME DISCOMFORT. AT THE PREVIOUS FOLLOW-UP, PERFORMED ON (B)(6) 2013 BATTERY IMPEDANCE WAS AT 1.53KOHMS. A REINTERVENTION WAS PERFORMED ON (B)(6) 2014 AND THE DEVICE WAS REPLACED. PHYSICIAN¿S COMMENT: VENTRICULAR OUTPUT WAS SET TO 3V/1MS ON THIS EXTRACTED PACEMAKER, BUT MODE WAS SAFER WITH LOW PACING %. CONSIDERING THESE CONDITIONS, IT COULD BE TOO EARLY TO REACH RRT. IN 5 YEARS AND 7 MONTHS THUS THE DEVICE WILL BE RETURNED FOR ANALYSIS.
REPORTEDLY, ON (B)(6) 2014, DURING A REGULAR FOLLOW-UP, THE PACEMAKER REACHED RECOMMENDED REPLACEMENT TIME (RRT) AND BATTERY IMPEDANCE WAS AT 10KOHMS. ACCORDING TO THE PATIENT, THE PATIENT FELT SOME DISCOMFORT. AT THE PREVIOUS FOLLOW-UP, PERFORMED ON (B)(6) 2013 BATTERY IMPEDANCE WAS AT 1.53KOHMS. A REINTERVENTION WAS PERFORMED ON (B)(6) 2014 AND THE DEVICE WAS REPLACED. PHYSICIAN'S COMMENT: VENTRICULAR OUTPUT WAS SET TO 3V/1MS ON THIS EXTRACTED PACEMAKER, BUT MODE WAS SAFER WITH LOW PACING PERCENTAGE. CONSIDERING THESE CONDITIONS, IT COULD BE TOO EARLY TO REACH RRT. IN 5 YEARS AND 7 MONTHS THUS THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341160 | REPLY | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | SORIN GROUP ITALIA S.R.L. - CRM FACILITY | REPLY DR | 2318 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |