FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 3863653
·
Received June 10, 2014
Report
- Report Number
- 2531779-2014-16266
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Report Date
- May 15, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP WAS OBSERVED TO HAVE A CRACK IN THE BATTERY COMPARTMENT FROM THE FINGER PAD TO THE PRIMARY CASE SEAL. THERE WAS NO EVIDENCE OF MOISTURE INSIDE THE PUMP. THERE WERE NO ISSUES WITH POWER TO THE PUMP DURING TESTING. THE BOLUS BUTTON WAS FOUND TO BE DETACHED FROM THE PUMP AND THE INTERNAL BUTTON PLUG WAS MISSING. NO INTERNAL CONTAMINATION WAS OBSERVED INSIDE THE BUTTON. (B)(6).
Description of Event or Problem · 1
THE PRODUCT WAS RETURNED TO ANIMAS. PRODUCT ANALYSIS EVALUATED THE PUMP AND FOUND THAT THE BATTERY COMPARTMENT WAS CRACKED AND THE BOLUS BUTTON INTERNAL PLUG WAS MISSING. THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION COMPLETED ON (B)(4) 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 337586 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |