FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3863653 · Received June 10, 2014

Report

Report Number
2531779-2014-16266
Event Type
Malfunction
Date Received
June 10, 2014
Report Date
May 15, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP WAS OBSERVED TO HAVE A CRACK IN THE BATTERY COMPARTMENT FROM THE FINGER PAD TO THE PRIMARY CASE SEAL. THERE WAS NO EVIDENCE OF MOISTURE INSIDE THE PUMP. THERE WERE NO ISSUES WITH POWER TO THE PUMP DURING TESTING. THE BOLUS BUTTON WAS FOUND TO BE DETACHED FROM THE PUMP AND THE INTERNAL BUTTON PLUG WAS MISSING. NO INTERNAL CONTAMINATION WAS OBSERVED INSIDE THE BUTTON. (B)(6).

Description of Event or Problem · 1

THE PRODUCT WAS RETURNED TO ANIMAS. PRODUCT ANALYSIS EVALUATED THE PUMP AND FOUND THAT THE BATTERY COMPARTMENT WAS CRACKED AND THE BOLUS BUTTON INTERNAL PLUG WAS MISSING. THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION COMPLETED ON (B)(4) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337586 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1