FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3863538 · Received June 10, 2014

Report

Report Number
1416980-2014-18493
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
March 29, 2014
Report Date
May 17, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS EVALUATED AND THE ISSUE WAS IDENTIFIED THROUGH REVIEW OF THE EVENT HISTORY LOGS. THE CAUSE WAS DETERMINED TO BE A USE ERROR OF INAPPROPRIATELY BYPASSING THE DRAIN, NOT INCLUDING THE INITIAL DRAIN. THE HOMECHOICE / HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE GIVES INSTRUCTIONS ON HOW TO BYPASS THE DRAIN PHASE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 22:37:20. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 1830ML, INDICATING THE HOME PATIENT (HP) DRAINED 1830ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337759 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1