FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3863537 · Received June 10, 2014

Report

Report Number
2531779-2014-16393
Event Type
Malfunction
Date Received
June 10, 2014
Report Date
May 23, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS (B)(4) 2014 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED. UNRELATED TO THIS ISSUE, TESTING REVEALED THE TIME AND DATE RESET TO FACTORY SETTINGS. THE PUMP WAS OPENED AND EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE CIRCUIT BOARD HAD FAILED. ADDITIONALLY, THE AUDIO BOLUS BUTTON COVER WAS DETACHED FROM THE PUMP. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DISPLAY FAILURE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338406 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 71 YR