FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3863442 · Received June 10, 2014

Report

Report Number
3008973940-2014-00199
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
March 17, 2014
Report Date
March 26, 2014
Manufacturer
MEDTRONIC SINGAPORE OPERATIONS
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND THE PROXIMAL CONDUCTOR OF THE LEAD BECAME EXTRINSICALLY DISTORTED DUE TO PULLING/STRETCHING/OVERSTRESS. IT WAS NOTED THAT THERE WAS BLOOD ON THE DISTAL CONDUCTOR OF THE LEAD AND IT WAS OBSTRUCTED, THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BODY TISSUE/FIBROTIC GROWTH, THE INNER AND OUTER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT, AND VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT.

Additional Manufacturer Narrative · 1

A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT TWO DIFFERENT LEADS WERE ATTEMPTED TO BE PLACED BUT UPON POSITIONING BOTH LEADS HAD HIGH THRESHOLDS. IT WAS ALSO REPORTED THAT THERE WERE HELIX RETRACTION DIFFICULTIES. ONE HELIX WOULD NOT RETRACT AND THE OTHER RETRACTED AFTER AN EXCESSIVE NUMBER OF TURNS. A TINED LEAD WAS IMPLANTED INSTEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341070 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MEDTRONIC SINGAPORE OPERATIONS 5076-58

Patients

Seq Age Sex Outcome Treatment
1