FDA Adverse Event Malfunction Summary report: N

CAPSURE Z NOVUS

MDR report key: 3863416 · Received June 10, 2014

Report

Report Number
2649622-2014-06732
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
March 21, 2014
Report Date
March 21, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P850089
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCTS: P1501DR IMPLANTABLE PULSE GENERATOR (IPG) IMPLANTED: 2005 (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCE MEASUREMENTS OF 2800 OHMS. IN ADDITION, THE RIGHT ATRIAL (RA) LEAD TREND SHOWS A GRADUAL INCREASE IN IMPEDANCE. THE RV AND RA LEADS CURRENTLY REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341180 CAPSURE Z NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5554-45

Patients

Seq Age Sex Outcome Treatment
1 00044 YR 507652 LEAD