FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3863394 · Received June 10, 2014

Report

Report Number
2649622-2014-06716
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 22, 2014
Report Date
April 22, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND; FULL LEAD RETURNED AND ANALYZED.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, THE RV (RIGHT VENTRICULAR) LEAD WAS INITIALLY POSITIONED WITH NORMAL TESTING RESULTS. THE ATRIAL LEAD WAS THEN PLACED. UPON RETESTING THE RV LEAD, THE THRESHOLD WENT VERY HIGH AND PACING IMPEDANCE ALSO BECAME HIGH, WITH POSSIBLE LEAD FRACTURE NOTED. REPOSITIONING OF THE RV LEAD WAS ATTEMPTED WITH SIMILAR RESULTS. THE RV LEAD WAS NOT USED AND ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339108 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI58

Patients

Seq Age Sex Outcome Treatment
1 00077 YR 5086MRI52 LEAD