CAPSUREFIX MRI SURESCAN
Report
- Report Number
- 2649622-2014-06716
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- April 22, 2014
- Report Date
- April 22, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
EVALUATION SUMMARY: ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND; FULL LEAD RETURNED AND ANALYZED.
CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT DURING IMPLANT ATTEMPT, THE RV (RIGHT VENTRICULAR) LEAD WAS INITIALLY POSITIONED WITH NORMAL TESTING RESULTS. THE ATRIAL LEAD WAS THEN PLACED. UPON RETESTING THE RV LEAD, THE THRESHOLD WENT VERY HIGH AND PACING IMPEDANCE ALSO BECAME HIGH, WITH POSSIBLE LEAD FRACTURE NOTED. REPOSITIONING OF THE RV LEAD WAS ATTEMPTED WITH SIMILAR RESULTS. THE RV LEAD WAS NOT USED AND ANOTHER LEAD WAS SUCCESSFULLY IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 339108 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00077 YR | 5086MRI52 LEAD |