FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3863367 · Received June 10, 2014

Report

Report Number
2649622-2014-06683
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 28, 2014
Report Date
March 28, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID D224TRK ICD, IMPLANTED: (B)(6) 2011; 5076-52 LEAD, IMPLANTED: (B)(6) 2003. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD HIGH PACING THRESHOLDS THAT DEPLETED THE DEVICE BATTERY IN THREE YEARS. THE PACE/SENSE PORTION OF THE LEAD WAS CAPPED AND REPLACED. THE HIGH VOLTAGE PORTION REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340900 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00071 YR Hospitalization| R 2187-85 LEAD