SPRINT QUATTRO SECURE
Report
- Report Number
- 2649622-2014-06688
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- April 4, 2014
- Report Date
- April 4, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE RIGHT VENTRICULAR LEAD INTEGRITY ALERT, AND INDICATED THE IMPEDANCE ON THE RV (RIGHT VENTRICULAR) PACING LEAD WAS BEYOND THE EXPECTED UPPER RANGE. ANALYSIS OF THE DEVICE MEMORY ALSO INDICATED OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER). THE ANALYST NOTED: (B)(6) 2014 LEAD INTEGRITY ALERT (LIA) ALERT FOR HIGH RATE NST EVENTS AND SIC COUNTS; ALSO TRIGGERING RV LEAD NOISE WARNING. BEGINNING THE WEEK OF (B)(6) 2014, RV PACING IMPEDANCE SPIKED FROM 456 OHMS TO 2375 OHMS. SENSING INTEGRITY COUNTS (B)(4) BEGINNING (B)(6) 2014.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCES AND NOISE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341438 | SPRINT QUATTRO SECURE | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6947 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00074 YR | Hospitalization| R |