FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3863317 · Received June 10, 2014

Report

Report Number
2649622-2014-06655
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 7, 2014
Report Date
April 7, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: PRODUCT ID 6949-65 DEFIB LEAD, IMPLANTED: (B)(6) 2005; 5076-52 CRDM NON-DEFIB LEAD, IMPLANTED: (B)(6) 2005. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT RECENT ONSET OF SHORT INTERVAL COUNTS (SIC) EVENTS WERE NOTED. PERFORMANCE DATA REGARDING THE LEAD REVEALED THAT VENTRICULAR -SIC WAS ONLY INCREMENTING DURING ATRIAL FIBRILLATION (AF) EPISODES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340857 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00058 YR Required Intervention D314DRG