FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE S
MDR report key: 3863302
·
Received June 10, 2014
Report
- Report Number
- 2649622-2014-06645
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- March 26, 2014
- Report Date
- March 26, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THE DISTAL LV (LOW VOLTAGE) ELECTRODE OF THE LEAD WAS COVERED IN BLOOD. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT AT THE IMPLANT PROCEDURE, THE HELIX ON THE LEAD WOULD NOT EXTEND. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341366 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 6935M62 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|