FDA Adverse Event Injury Summary report: N

REVO MRI SURESCAN

MDR report key: 3863298 · Received June 10, 2014

Report

Report Number
9614453-2014-01342
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 9, 2014
Report Date
April 9, 2014
Manufacturer
IPG MFG SWITZERLAND
Product Code
LWP
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: A 5086MRI52 LEAD: IMPLANTED: (B)(6) 2011. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED A PARTIAL ELECTRICAL RESET. CRITICAL RAM PARITY ERROR OCCURRED IN ADDRESS 17 14 ON (B)(6) 2014. POWER ON RESET (POR) SEVERITY IS LOW SO THE DEVICE SHOULD BE ABLE TO FULLY RECOVER AFTER RESET.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXPERIENCED A PARTIAL DEVICE RESET. THE RESET WAS CLEARED AND THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338935 REVO MRI SURESCAN PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE LWP IPG MFG SWITZERLAND RVDR01

Patients

Seq Age Sex Outcome Treatment
1 00067 YR Required Intervention 5086MRI58 LEAD