REVO MRI SURESCAN
Report
- Report Number
- 3004209178-2014-10996
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- April 15, 2014
- Report Date
- April 24, 2014
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- LWP
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 5086MRI52 LEAD IMPLANTED: (B)(6) 2011. (B)(4).
PRODUCT EVENT SUMMARY: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED A PARTIAL ELECTRICAL RESET. CRITICAL RAM PARITY ERROR OCCURRED IN ADDRESS 00 B2 ON 2014-(B)(6). POWER ON RESET (POR) SEVERITY IS LOW SO THE DEVICE SHOULD BE ABLE TO FULLY RECOVER AFTER RESET. PATIENT IS UNDERGOING RADIATION THERAPY.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXPERIENCED AN ELECTRICAL RESET. THE RESET WAS CLEARED AND THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 339204 | REVO MRI SURESCAN | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | RVDR01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00080 YR | Required Intervention | 5086MRI58 LEAD |