FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3863244 · Received June 10, 2014

Report

Report Number
2649622-2014-06611
Event Type
Injury
Date Received
June 10, 2014
Report Date
March 18, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 693565 LEAD, IMPLANTED: (B)(6) 2012. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS DIAGNOSED WITH SEPSIS AND THE PRIMARY SOURCE OF THE INFECTION WAS REPORTED TO BE THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM. BLOOD CULTURES REVEALED A (B)(6) INFECTION. THE ICD SYSTEM WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341261 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-52

Patients

Seq Age Sex Outcome Treatment
1 00085 YR Hospitalization| L| R D284DRG ICD