SPRINT QUATTRO SECURE

MDR report key: 3863238 · Received June 10, 2014

Report

Report Number
2649622-2014-06616
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 17, 2014
Report Date
April 17, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Removal / Correction Number
Z-0475-2011
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING DUE TO NON-PHYSIOLOGIC SIGNALS/SIC (SENSING INTEGRITY COUNTER) AND A LEAD IMPEDANCE OUT OF RANGE ALERT. (B)(4) LIFETIME V-SIC ARE RECORDED BEGINNING 21 APR 2014. AN OUT OF TOLERANCE (OOT) SUBTHRESHOLD LEAD IMPEDANCE ALERT OCCURRED ON (B)(6) 2014. THERE ARE 2 NON-SUSTAINED TACHYCARDIA (NST) EVENTS OF LESS THAN 220 MS V-V CYCLE RECORDED BEGINNING (B)(6) 2014.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS DEVICE WAS INCLUDED IN THAT FIELD ACTION.  BASED ON THE INFORMATION RECEIVED AND WITHOUT THE RETURN OF THE PRODUCT, IT COULD NOT DETERMINE THIS DEVICE PERFORMED AS DESCRIBED IN THE FIELD ACTION. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED AN INAPPROPRIATE THERAPY. IT WAS NOTED THE PATIENT HAS A SMALL R WAVE AMPLITUDE WITH T-WAVE OVERSENSING (TWOS). THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED AN INAPPROPRIATE THERAPY. IT WAS NOTED THE PATIENT HAS A SMALL R WAVE AMPLITUDE WITH T-WAVE OVERSENSING (TWOS). THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341259 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00053 YR Hospitalization| L| R D284VRC ICD