SPRINT QUATTRO SECURE S
Report
- Report Number
- 2649622-2014-06586
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 7, 2014
- Report Date
- March 14, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND NO ANOMALIES WERE FOUND. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THIS EVENT OCCURRED OUTSIDE THE U.S. WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
IT WAS REPORTED THAT A FEW DAYS AFTER IMPLANT THE RIGHT VENTRICULAR (RV) LEAD THRESHOLD INCREASED TO A HIGH THRESHOLD. NOISE WAS ALSO NOTED ON THE RV LEAD WHICH TRIGGERED A LEAD INTEGRITY ALERT. THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341193 | SPRINT QUATTRO SECURE S | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 693565 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |