CAPSUREFIX MRI SURESCAN
Report
- Report Number
- 2649622-2014-06563
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- March 28, 2014
- Report Date
- March 28, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5086MRI52 LEAD, IMPLANTED: (B)(6) 2013. (B)(4).
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE ANALYSIS REVEALED THE INNER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT. THE OUTER INSULATION OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION AND THE VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. THE ANALYST NOTED THAT THE BREACHED INSULATION DID CONTRIBUTE TO THE ELECTRICAL COMPLAINT OF UNDERSENSING AND IT APPEARS TO HAVE BEEN CAUSED AT THE IMPLANT.
IT WAS REPORTED THAT THE PATIENT HAD EXTRACARDIAC MUSCLE STIMULATION AND THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING WITH NO MEASURABLE P WAVES. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 339539 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI45 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00043 YR | Hospitalization| R | RVDR01 IPG |