FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3863150 · Received June 10, 2014

Report

Report Number
2649622-2014-06563
Event Type
Injury
Date Received
June 10, 2014
Date of Event
March 28, 2014
Report Date
March 28, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT MEDICAL PRODUCTS: 5086MRI52 LEAD, IMPLANTED: (B)(6) 2013. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED AND ANALYZED. THE ANALYSIS REVEALED THE INNER INSULATION OF THE LEAD WAS EXTRINSICALLY BREACHED DUE TO A CUT. THE OUTER INSULATION OF THE LEAD WAS OBSERVED TO HAVE BLOOD INGRESSION AND THE VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. THE ANALYST NOTED THAT THE BREACHED INSULATION DID CONTRIBUTE TO THE ELECTRICAL COMPLAINT OF UNDERSENSING AND IT APPEARS TO HAVE BEEN CAUSED AT THE IMPLANT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD EXTRACARDIAC MUSCLE STIMULATION AND THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING WITH NO MEASURABLE P WAVES. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
339539 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI45

Patients

Seq Age Sex Outcome Treatment
1 00043 YR Hospitalization| R RVDR01 IPG