CAPSURE SENSE
Report
- Report Number
- 2649622-2014-06524
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- December 30, 2013
- Report Date
- December 30, 2013
- Manufacturer
- MPRI
- Product Code
- DTB
- PMA / PMN Number
- P830061
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: 407452, LEAD, IMPLANTED: (B)(6) 2005. (B)(4).
PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND. THE VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED APPARENT EXPLANT DAMAGE.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THERE WAS ELEVATED THRESHOLDS OBSERVED IN THE RIGHT VENTRICULAR (RV) LEAD AND THE LEAD WAS EXPLANTED AND REPLACED. IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL (RA) LEAD "CAME OUT" DURING THE PROCEDURE. THE RA LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 338634 | CAPSURE SENSE | ELECTRODE, PACEMAKER, PERMANENT | DTB | MPRI | 457445 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00075 YR | Hospitalization| R | E2DR01AA IPG |