FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ 3.13 DUAL CHA
MDR report key: 3862968
·
Received February 20, 2014
Report
- Report Number
- 9615050-2014-01377
- Event Type
- Malfunction
- Date Received
- February 20, 2014
- Date of Event
- January 14, 2014
- Report Date
- January 22, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND THAT THE DEVICE ALARMED WITH A S233 (OVERTEMPERATURE-PSC) MALFUNCTION ALARM CODE AND WAS NOTED IN THE DEVICE HISTORY. FURTHER TESTING FOUND THAT THE DEVICE FAN WAS ROTATING TOO SLOWLY TO COOL THE DEVICE. THE PROBABLE CAUSE WAS A BROKEN FAN. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT THE DEVICE ALARMED WITH A S233 (OVER TEMPERATURE-PSC) MALFUNCTION ALARM CODE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 106734 | SYMBIQ 3.13 DUAL CHA | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |