FDA Adverse Event Malfunction Summary report: N

SYMBIQ 3.13 DUAL CHA

MDR report key: 3862968 · Received February 20, 2014

Report

Report Number
9615050-2014-01377
Event Type
Malfunction
Date Received
February 20, 2014
Date of Event
January 14, 2014
Report Date
January 22, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THAT THE DEVICE ALARMED WITH A S233 (OVERTEMPERATURE-PSC) MALFUNCTION ALARM CODE AND WAS NOTED IN THE DEVICE HISTORY. FURTHER TESTING FOUND THAT THE DEVICE FAN WAS ROTATING TOO SLOWLY TO COOL THE DEVICE. THE PROBABLE CAUSE WAS A BROKEN FAN. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE DEVICE ALARMED WITH A S233 (OVER TEMPERATURE-PSC) MALFUNCTION ALARM CODE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
106734 SYMBIQ 3.13 DUAL CHA 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA