FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3862890 · Received February 20, 2014

Report

Report Number
1720753-2014-01683
Event Type
Malfunction
Date Received
February 20, 2014
Date of Event
February 10, 2014
Report Date
February 20, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE VIDEO CONTROLLER AND IMAGE PROCESSOR BOARDS WERE RESEATED DURING SERVICE. THE SYSTEM WAS TESTED AND FOUND TO WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEY WERE UNABLE TO VIEW FLUOROSCOPIC IMAGE BECAUSE THE SYSTEM DISPLAYED A BLACK IMAGE. THIS ISSUE WILL CAUSE THE SYSTEM TO BE UNUSABLE DUE TO THE INABILITY TO SEE THE LIVE IMAGE. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
106763 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1