INRATIO2
Report
- Report Number
- 2027969-2014-00139
- Event Type
- Malfunction
- Date Received
- February 19, 2014
- Date of Event
- January 27, 2014
- Report Date
- January 27, 2014
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K092987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NOT APPLICABLE
Narratives
IT IS INDICATED THAT PRODUCT IS NOT RETURNING FOR EVALUATION. THEREFORE, INVESTIGATION OF THE COMPLAINT TO DETERMINE ROOT CAUSE CANNOT BE COMPLETED. RETAIN STRIP TESTING RESULTS MET BOTH ACCURACY AND REPEATABILITY CRITERIA. A REVIEW OF THE BATCH RECORD FOR THE LOT DID NOT UNCOVER ANY NON-CONFORMANCES. RESULTS OF THE QUERY OF SIMILAR INCIDENT IN THE COMPLAINT SYSTEM FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. ALTHOUGH THE ROOT CAUSE ANALYSIS DID NOT INCLUDE RETURNED TESTING, IMPROPER TECHNIQUES WERE IDENTIFIED IN THE COMPLAINT. THESE COULD NOT BE RULED OUT AS A CAUSE OF THE UNEXPECTED RESULTS. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME. THE ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING.
CALLER ALLEGED DISCREPANT INRATIO2 RESULTS. RESULTS AS FOLLOWS: INRATIO2: 1.6, RE-TEST: 2.7. TIME BETWEEN TESTS: < 2 MINUTES. THERAPEUTIC RANGE: 2.5-3.5. PATIENT SELF TESTER REPORTS THEY PERFORMED 2 FINGER STICKS ON THE SAME FINGER. CUSTOMER WAS UNABLE TO OBTAIN A LARGE ENOUGH DROP TO TEST ON THE FIRST FINGER STICK. THEY THEN PERFORMED A NEW FINGER STICK, TESTED AND RECEIVED A 1.6. THEY WERE CONCERNED WITH THAT READING SO THEY CLEANED THE FINGER AND SQUEEZED OUT A SECOND SAMPLE USING THE SAME FINGER STICK, INR= 2.7.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 104349 | INRATIO2 | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100071 | 329655 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | COUMADIN |