FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3862671 · Received June 10, 2014

Report

Report Number
2649622-2014-06791
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 1, 2014
Report Date
April 1, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED. ANALYSIS WAS PERFORMED AND NO ANOMALIES WERE FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT PROCEDURE, IT WAS IMPOSSIBLE TO FIX THE LEAD TO THE VENTRICULAR WALL IN SPITE OF THE SEVERAL ATTEMPTS. THE FLUOROSCOPY SHOWED THAT THE HELIX WAS NOT COMPLETELY JOINED WHEN THE PHYSICIAN WAS TRYING TO RELOCATE THE LEAD. THE LEAD WAS NOT IMPLANTED AND ANOTHER LEAD WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340856 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00028 YR