FDA Adverse Event Malfunction Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3862669 · Received June 10, 2014

Report

Report Number
2649622-2014-06790
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
April 28, 2014
Report Date
April 28, 2014
Manufacturer
MPRI
Product Code
NVY
PMA / PMN Number
P920015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE ACTUAL PRODUCT WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA WAS COLLECTED FROM THE DEVICE AND WAS ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED THE IMPEDANCE ON THE SUPERIOR VENA CAVA (SVC) DEFIBRILLATION COIL WAS BEYOND THE EXPECTED UPPER RANGE. ALERT FOR OUT OF RANGE SUBTHRESHOLD LEAD IMPEDANCE ALERT ON (B)(6) 2014. SVC LEAD IMPEDANCE WITHIN NORMAL LIMITS; BEGINS TRENDING OUT OF RANGE WEEK ENDING (B)(6) 2014.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCTS: 407645 LEAD, IMPLANTED 2006 (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD SHOWED A JUMP IN SUPERIOR VENA CAVA (SVC) COIL IMPEDANCE TO HIGH VALUES THEN IT IMPEDANCE CAME BACK DOWN. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
338957 SPRINT QUATTRO SECURE DRUG-ELUTING PERMANENT DEFIBRILLATOR ELECTRODES NVY MPRI 6947M62

Patients

Seq Age Sex Outcome Treatment
1 00089 YR (B)(4) ICD