FDA Adverse Event Injury Summary report: N

CAPSUREFIX MRI SURESCAN

MDR report key: 3862587 · Received June 10, 2014

Report

Report Number
2649622-2014-06832
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 16, 2014
Report Date
April 16, 2014
Manufacturer
MPRI
Product Code
NVN
PMA / PMN Number
P090013
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE ACTUAL LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING ASSOCIATED WITH THE RIGHT VENTRICULAR LEAD.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLD AND OVERSENSING RESULTING FROM A SUSPECTED INSULATION FAILURE. THE RV LEAD WAS CAPPED AND REPLACED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
341032 CAPSUREFIX MRI SURESCAN DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN MPRI 5086MRI52 LFP114383V

Patients

Seq Age Sex Outcome Treatment
1 00075 YR Hospitalization| R A3DR01 IPG