CAPSUREFIX MRI SURESCAN
Report
- Report Number
- 2649622-2014-06832
- Event Type
- Injury
- Date Received
- June 10, 2014
- Date of Event
- April 16, 2014
- Report Date
- April 16, 2014
- Manufacturer
- MPRI
- Product Code
- NVN
- PMA / PMN Number
- P090013
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
Narratives
PRODUCT EVENT SUMMARY: THE ACTUAL LEAD WAS NOT RETURNED FOR ANALYSIS. HOWEVER, PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS RECEIVED AND ANALYZED. ANALYSIS OF THE DEVICE MEMORY INDICATED OVERSENSING ASSOCIATED WITH THE RIGHT VENTRICULAR LEAD.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLD AND OVERSENSING RESULTING FROM A SUSPECTED INSULATION FAILURE. THE RV LEAD WAS CAPPED AND REPLACED SUCCESSFULLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 341032 | CAPSUREFIX MRI SURESCAN | DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE | NVN | MPRI | 5086MRI52 | LFP114383V |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00075 YR | Hospitalization| R | A3DR01 IPG |