FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3862581 · Received January 23, 2014

Report

Report Number
1314492-2014-04352
Event Type
Malfunction
Date Received
January 23, 2014
Date of Event
November 1, 2013
Report Date
December 30, 2013
Manufacturer
BAXTER HLTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED AND EVALUATED BY BAXTER. THIS COMPLAINT WAS OPENED DURING THE INVESTIGATION OF COMPLAINT (B)(4). THE IDENTIFIED "AIR-IN-LINE" ALARMS WERE CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADD'L INFO IS RECEIVED A FOLLOW-UP WILL BE SENT. THE ULTRASONIC SENSOR, ULTRASONIC PUSHER AND PROCESSOR PRINTED CIRCUIT BOARD WERE REPLACED.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED SPECTRUM INFUSION PUMP, 17 OCCURRENCES OF "AIR-IN-LINE" ALARMS WERE IDENTIFIED IN THE EVENT HISTORY LOG. ANY PATIENT INJURY OR MEDICAL INTERVENTION IS UNK SINCE THESE ITEMS WERE FOUND DURING EVALUATION OF THE DEVICE. NO ADD'L INFO IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55302 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HLTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1