FDA Adverse Event Malfunction Summary report: N

CAPSUREFIX NOVUS

MDR report key: 3862561 · Received June 10, 2014

Report

Report Number
2649622-2014-06826
Event Type
Malfunction
Date Received
June 10, 2014
Date of Event
March 28, 2014
Report Date
March 28, 2014
Manufacturer
MPRI
Product Code
DTB
PMA / PMN Number
P930039
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: THE FULL LEAD WAS RETURNED, ANALYZED, AND THERE WAS BLOOD ON THE DISTAL CONDUCTOR OF THE LEAD AND IT WAS OBSTRUCTED. VISUAL SUMMARY ANALYSIS OF THE LEAD INDICATED DAMAGE AT IMPLANT. THE ANALYST NOTED THAT THE STYLET INSERTION TEST FAILED. DESTRUCTIVE ANALYSIS WAS PERFORMED AND VISUAL INSPECTION OF CONDUCTOR AND FOUND A BLOOD OBSTRUCTION.

Additional Manufacturer Narrative · 1

CORRECTED INFORMATION: NO EVAL EXPLAIN CODE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT OF THE RIGHT VENTRICULAR (RV) LEAD THERE WAS PLACEMENT DIFFICULTY BECAUSE, THE LEAD WAS NOT LONG ENOUGH DUE TO PATIENT ANATOMY. IT WAS ALSO REPORTED THAT THE STYLET WAS STICKING IN THE LEAD AFTER MULTIPLE POSITIONING ATTEMPTS. THE LEAD WAS REMOVED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340973 CAPSUREFIX NOVUS ELECTRODE, PACEMAKER, PERMANENT DTB MPRI 5076-58

Patients

Seq Age Sex Outcome Treatment
1 00080 YR