FDA Adverse Event
Malfunction
Summary report: N
SPRINT QUATTRO SECURE
MDR report key: 3862549
·
Received June 10, 2014
Report
- Report Number
- 2649622-2014-06856
- Event Type
- Malfunction
- Date Received
- June 10, 2014
- Date of Event
- March 26, 2014
- Report Date
- March 26, 2014
- Manufacturer
- MPRI
- Product Code
- LWS
- PMA / PMN Number
- P920015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCTS: 5076 LEAD IMPLANTED: 2013 (B)(6). (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT HAD A RECENT DEVICE REPLACEMENT PROCEDURE AND SUBSEQUENTLY AN ALERT WAS TRIGGERED FOR HIGH LEAD IMPEDANCE IN THE HIGH VOLTAGE COILS. A REAL TIME MEASUREMENT REVEALED IMPEDANCE VALUES WITHIN THE EXPECTED RANGE AND IT WAS INDICATED AS AN INTERMITTENT ISSUE. THE RIGHT VENTRICULAR (RV) LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 340970 | SPRINT QUATTRO SECURE | DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER | LWS | MPRI | 694758 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00081 YR | (B)(4) ICD |