FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO SECURE

MDR report key: 3862482 · Received June 10, 2014

Report

Report Number
2649622-2014-06873
Event Type
Injury
Date Received
June 10, 2014
Date of Event
April 13, 2014
Report Date
April 25, 2014
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY : THE FULL LEAD WAS RETURNED IN SEGMENTS AND ANALYZED; ANALYSIS REVEALED A FLEX FRACTURE OF THE DISTAL CONDUCTOR.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED INAPPROPRIATE SHOCKS DUE TO THE RIGHT VENTRICULAR LEAD HAVING "PARASITE POTENTIALS" (ARTIFACT), WHICH INDICATED ELECTRODE DAMAGE. THE SENSING INTEGRITY COUNTER WAS ELEVATED AND NON-SUSTAINED TACHYCARDIA EPISODES WERE NOTED. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
340471 SPRINT QUATTRO SECURE DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694765

Patients

Seq Age Sex Outcome Treatment
1 00046 YR Hospitalization| L| R